Have a Question?
Stitching Considerations especially for Scoliosis Cases
Why Stitching Can Change Measurements: A Step-by-Step Review
Video Link: https://youtu.be/cvSaAmT3fAU?si=Tnc8zRHBUuLetlYe
1. Understand why stitched scoliosis values can change 0:01

- This tutorial is for advanced users stitching scoliosis sectionals.
- The main point: measurement changes after stitching are often expected, not necessarily a software bug.
- The changes usually come from how the x-rays were taken, chosen, and aligned.
2. Remember that each sectional is shot from a different central ray angle 0:52

- Sectionals are usually taken as multiple x-rays.
- Each image has a different central ray angle, so the same vertebra can look slightly different from one film to another.
- That changes the perceived end plates and Cobb line placement, which can alter the measured angle.
3. Prefer stitching to a new case when possible 1:40

- For scoliosis, it is often easier to stitch into a new case rather than an existing one.
- This is especially helpful when the curve extends beyond the edges of the film.
- A new case makes it easier to:
- work with the full spine,
- identify the true start/end of the curve,
- compare full-spine studies later,
- avoid confusion from re-editing an existing case.
4. Set patient expectations before stitching 3:19

- Tell the patient up front that the x-rays are being stitched together.
- Explain that small measurement changes can happen because the goal is to create an overall picture of spinal alignment.
- You do not need to explain the physics in detail—just enough so the patient understands why values may shift.
5. Fine-tuning points after stitching is normal 2:59

- Post-processing often places the images in the general vicinity.
- The doctor may then adjust digitization points or even the scoliosis level after stitching.
- That is acceptable, but it reinforces why the clinician must understand how projection affects the result.
6. Choose the correct film for the overlay 4:34

- A major source of change is choosing the wrong x-ray as the top image during stitching.
- In the example, the lumbar film was placed on top instead of the thoracic film.
- Because the overlap and visible anatomy were different, the Cobb angle changed significantly even though the points were not moved much.
7. Understand why Cobb angles are especially sensitive 6:13

- Cobb angles are still the standard and must be reported.
- But Cobb measurements are highly sensitive to small endpoint changes.
- That means even a slight difference in projection or endpoint selection can create a large change in the final angle.
8. Compare Cobb with other methods to see the difference 6:52

- In the example, the Risser-Ferguson measurement changed only slightly.
- The Cobb angle changed much more dramatically.
- This shows that Cobb is more vulnerable to projection and endpoint differences than some other methods.
9. Restitch using the better overlay image 7:18

- The fix in the example was to restitch without moving the points, but choose the thoracic film as the top overlay.
- Once the correct film was used, the measurements returned to nearly the original values.
- This confirms the issue was image selection and projection, not a software error.
10. Know the main reasons stitched values change 8:08

- The clinician may move points during post-processing.
- The wrong x-ray may be chosen as the overlay.
- The central ray differs between sectionals, so the same vertebra is viewed from a different angle.
- All of these can cause small or large measurement differences after stitching.
11. Use full-spine AP imaging when available, but know its limits 9:00

- A full-spine AP image reduces some projection distortion because it captures the spine in one shot.
- But sectionals can still be useful because they reduce distortion in the areas they cover.
- The key is understanding that different imaging methods produce different perspectives.
12. Watch for upcoming stitching improvements 10:03

- Future updates to the stitching module will make it easier to see ghosted overlays, lines, and digitization points.
- That should help users judge alignment more accurately during stitching.
- Version 27 and later will also improve how scoliosis is handled in reporting.
13. Use scoliosis digitization more directly in newer versions 10:42

- In older versions, users often had to digitize with the 3-point modified Risser-Ferguson method, then restitch and redo as scoliosis.
- In newer versions, you can digitize directly as scoliosis.
- This saves time and still reports the Cobb angle appropriately without incorrectly labeling every small curve as a scoliosis diagnosis.
14. The same principles apply on lateral views 11:48

- Lateral stitching can also produce changes, though usually less than AP views.
- That is because the lateral view looks more along the axis of rotation, with cleaner end plates and vertebral bodies.
- Still, projection differences can happen, so the same caution applies.
15. Final takeaway 12:22

- If stitched scoliosis values change, that does not automatically mean something is wrong.
- The likely causes are:
- projection differences,
- image selection,
- point placement,
- and the sensitivity of Cobb measurements.
- The clinician should understand this well enough to explain it to the patient and to choose the best stitching workflow going forward.